RESIDUAL OZONE ANALYZER

HydroACT Series

The HydroACT Residual Ozone Analyzer utilizes the very latest and best ozone sensors available in the world today.  The membrane-covered sensor is insensitive to changing pH and interference from other oxidants, uses no reagents, is extremely stable, and has a reduced maintenance requirement compared to most other Ozone sensors.

  • Video Testimonial: Chlorine Monitoring in Distribution

    Having accurate, reliable, and real-time chlorine monitoring, both in the water treatment plant, and in the distribution system, is critical. Check out this two-minute customer testimonial video profiling a municipality that decided to use Chemtrac chlorine analyzers throughout their distribution system. This was only after the operations personnel first experienced the reliability of the units that were installed in the plant, and then realized over time just how easy the analyzers were to maintain. 

  • Description

    The HydroACT Residual Ozone Analyzer utilizes the very latest and best ozone sensors available in the world today. The membrane covered sensor is insensitive to changing pH and interference from other oxidants, uses no reagents, is extremely stable, and has a reduced maintenance requirement compared to most other Ozone sensors. HydroACT offers a versatile solution to your monitoring needs with the flexibility to accept additional sensors, as well as offering expandable I/O and PID control loops.

    Standard Features

    • Includes analyzer, ozone sensor, flow cell, 1 analog output, 1 year consumables
    • No moving parts, no reagents
    • Event logging

    Ozone Dosing Application

    • Expandable sensor inputs and I/O
    • Color display and data-logging with microSD card download
    • Modbus or PROFIBUS communictions
    • Chemical feed control capability (e.g. PID)

    Benefits

    • Stable and reliable – excellent process control
    • Suitable for all potable, process and salt waters
    • No reagents or moving parts
    • Up to 6 months between maintenance
    • Up to 3 months between calibration
    • Resistant to detergents in the water

  • Specifications

    Type  Membrane-covered amperometric two-electrode system
    Measured  Residual Ozone
    Probe Ranges 0.05 – 0.2, 0.05 – 2, 0.05 – 5, or 0.05 – 10 mg/L (ppm)
    Resolution  0.01 mg/L (ppm)
    Reproducibility ±5%
    Working Electrode  Gold cathode
    Counter Electrode Silver/silver halide
    Membrane Material  Micro-porous hydrophillic membrane
    Flow Rate  15 to 60 L/hr.
    Stability  -1% per month,  (without calibration)
    Temperature Compensation  Automatically by integrated thermistor (ATC)
    Temperature Range  > 0 up to 50° C
    pH Range  pH 2 – pH 11
    Permissible Over-Pressure  7.25 psi (0.5 bar)
    First Polarization Time  120 min.
    Re-Polarization Time  30 min.
    Zero-Point Adjustment  Not necessary
    Calibration  Manual using a suitable ozone test kit
    Housing Material  PVC, silicone, polycarbonate, stainless steel
    Dimensions  Diameter approx. 0.98 in., length 6.89 in.
    Replacement Intervals
         Membrane Annually
         Electrolyte  Every 3 to 6 months